Note: All specifications apply to operation at 77°F (25°C)
| Display | Five Digits Top Display (-9999 to 99999)
 | ¾” (17.8 mm) high 14-segment alphanumeric
 Automatic lead zero blanking
 | 
| Seven Characters (Engineering Units) | ½” (10.2 mm) high 14-segment alphanumeric
 | 
| Symbols | High & Low Alarm, Password Lock | 
| Backlight | White | 
| Display Update Rate | Ambient > -13°F: 2 Updates/Second Ambient < -13°F: 1 Update/5 Seconds
 | 
| Overrange | Display flashes 99999 | 
| Underrange | Display flashes -9999 | 
| Programming Method | Four through-window buttons when cover is installed Four internal push-buttons when cover is removed
 | 
| Recalibration | Recalibration is recommended at least every 12 months. | 
| Max/Min Display | Max/Min readings reached by the process are stored until reset by the user or until power to the meter is turned off. | 
| Password | Programmable password restricts modification of programmed settings | 
| Non-Votalite Memory | All programmed settings are stored in non-volatile memory for a minimum
of ten years | 
| Normal Mode Rejection | 64 dB at 50/60 Hz | 
| Accuracy | ±0.05% FS ±0.001mA | 
| Temperature Range | Operating Temperature: -40°F to 167°F(-40°C to 75°C) for safe area products
 -40°F to 158°F (-40 to 70°C) for hazardous area products
 Storage Temperature: -40°F to 185°F (-40 to 85°C)
 | 
| Relative Humidity | 0 to 90% non-condensing Printed circuit boards are conformally coated
 |